Diagnostics
Abstract
For the last few decades, the immunochromatographic assay has been used for the rapid detection of biological markers in infectious diseases in humans and animals The assay, also known as lateral flow assay, is utilized for the detection of antigen or antibody in human infectious diseases. There are a series of steps involved in the development of these immuno-chromatographic test kits, from gold nano colloids preparation to nitrocellulose membrane coating (NCM). These tests are mostly used for qualitative assays by a visual interpretation of results. For the interpretation of the results, the color intensity of the test zone is therefore very significant. Herein, the study was performed on a malaria antigen test kit. Several studies have reported the use of gold nanoparticles (AuNPs) with varying diameters and its binding with various concentrations of protein in order to optimize tests. However, none of these studies have reported how to fix (improve) test zone band intensity (color), if different sized AuNPs were synthesized during a reaction and when conjugated equally with same amount of protein. Herein, different AuNPs with average diameter ranging from 10 nm to 50 nm were prepared and conjugated equally with protein concentration of 150 µg/mL with KD = 1.0 × 10-3. Afterwards, the developed kits' test zone band intensity for all different sizes AuNPs was fixed to the same band level (high) by utilization of an ultraviolet-visible spectrophotometer. The study found that the same optical density (OD) has the same test zone band intensity irrespective of AuNP size. This study also illustrates the use of absorption maxima (λ max) techniques to characterize AuNPs and to prevent wastage of protein while developing immunochromatographic test kits.
DOI
10.3390/diagnostics10110880
Publication Date
10-29-2020
Keywords
UV/Vis spectrophotometer, gold nanoparticles sensor, immuno-chromatographic, lateral flow assay, malaria pan rapid diagnostic strip, point-of-care
ISSN
2075-4418
Recommended Citation
Manta P, Nagraik R, Sharma A, Kumar A, Verma P, Paswan SK, Bokov DO, Shaikh JD, Kaur R, Leite AV, Filho SB, Shiwalkar N, Persaud P, Kapoor DN. Optical Density Optimization of Malaria Pan Rapid Diagnostic Test Strips for Improved Test Zone Band Intensity. Diagnostics. 2020; 10(11). doi: 10.3390/diagnostics10110880.